CM-8822 Digital Coating Thickness Gauge 0~40mil
Spec
The F probes measure the thickness of non-magnetic materials on magnetic materials. The N probes measure the thickness of non-conductive coatings on non-magnetic metals.
Features
* Two measuring methods: magnetic induction (F) and eddy current (N)
* Two measurement mode: Single and Continuous(Notice:Continuous mode is convenient to measure the coating thickness of curved objects and tiny objects.)
* Metric/imperial system selectable.
* Manual or automatic shut down.
* Automatic memory of calibration value,and automatic recognize the substrate.
* Operation process have buzzing reminder,while no buzzing in continuous mode.
Principle
Principles |
Applications |
Examples |
F Type
Magnetic Induction
(Iron Base) |
Measure the thickness of non-magnetic materials on magnetic materials |
Galvanizing layer, lacquer layer, porcelain enamel layer, phosphide layer, copper tile, aluminium tile, some alloy tile, paper etc |
NF Type
Eddy Current
(Aluminium Base) |
Measure the thickness of non-conductive coatings on non-magnetic metals |
Anodizing, varnish, paint, enamel, plastic coatings, powder, etc. Applied to aluminum, brass, non-magnetic stainless steel, etc |
Specifications
Principle |
F Magnetic Induction & NF Type Eddy Current |
Range |
0~1000µm/0~40mil |
Resolution |
0.1µm/1µm |
Accuracy |
±1~3%n or ±2.5µm |
Min.Radius Workpiece |
F Type:convex 1.5mm/concave 25mm |
NF Type:convex 3mm/concave 50mm |
Min.Measuring Area |
6mm |
Min.Sample Thickness |
0.3mm |
Metric/Imperial |
Convertible |
Battery Indicator |
Low Battery Indicator |
Automatic Power Off |
√ |
Operating Conditions |
Temperature:0~50ºC |
Humidity:<80%RH |
Power Supply |
4x1.5V AA (UM-3)Battery |
Dimensions |
161x69x32mm |
Weight |
250g(Not Including Batteries) |
Accessories
Main Unit
Probe (F Type)
Probe (NF Type)
Calibration Base (F)
Calibration Base (NF)
Calibration Foils (1 set, 4 pieces)
Carrying Case (B04)
Operation Manual